Manual control mode, automatic control mode, sequence control mode, automatic segmentation remotely control, acceptable correlation control.
The light intensity and time can be independently adjusted, controlled and set for each culture test tube.can be using light sources of different colors
Improved stirring and gas distribution system
Integrated sensors and monitoring systems
Modular and scalable design
Compatibility and versatility
Easy to clean and disinfect
Manual control mode, automatic control mode, sequence control mode, automatic segmentation remotely control, acceptable correlation control.
can feeding, bubble enemy, such as measurement of pH, DO.
show that all the parameters of the trend curve analysis.store, display, analyze all the historical curve parameters,display all the parameters of the batch report.
PH, DO and other parameters of control and associated controls can be adjusted.
Name | Volume (Customize) | D/L ratio |
Glass Bioreactor Fermenter (Customize) | 0.5L | 1:2/1:2.5 |
1L | 1:2/1:2.5 | |
3L | 1:2/1:2.5 | |
5L | 1:2/1:2.5 | |
10L | 1:2/1:2.5 | |
15L | 1:2/1:2.5 | |
Tank | Single glass vessel with stainless steel bottom. Lid made by stainless steel 316L with ports, condenser on the cover | |
Lighting system | external stainless steel light source, intensity manually adjustable. The light source system is detachable, making it easy to install light sources of different colors; the light intensity is between 0-3000 umol/m2/s; the light intensity and time can be independently adjusted, controlled and set for each culture test tube. | |
Stirring system | Mechanical drive or Magnetic drive (bottom/top drive), 3 six-blade paddle and 1 foam breaker, each paddle height adjustable.Stir speed: 50~1000rpm/50-500rpm | |
Sterilization | Off-site sterilization (Autoclave) | |
Gas control | Control by rotor meter, air flow 2vvm. | |
Feeding | Automatically feeding by peristaltic pump | |
Anti-foam control | Automatically add antifoamer by peristaltic pump | |
Temperature control | Auto-control, electric heating range water source add 5℃ to 70℃ | |
PH control | PH 2.00-12.00±0.1, auto-control adding acid/base by peristaltic pump, pH sensor from Hamilton/Mettler | |
DO control | 0-150±3% displaying precision 0.1%, DO sensor from Hamilton/Mettler, DO coupling with agitation, oxygen. | |
Sampling | Manual sampling | |
Control parameters | 10 or 15 inch touch screen , Siemens PLC with control parameters (stirring, pH, DO, temperature, feeding, alarm, gas etc.), data analysis, data record, data output, language(RU/EN/CN) and other customized functions | |
Optional configuration | Remote control, Mass flow meter(auto control gas), O2/N2/CO2, gas(CO2) exhaust analysis, Level sensor, Turbidity(OD) sensor, ORP sensor, weight system, Live cell sensor(Biomass), Printing, Siemens touch screen, etc., |